Products

WS-400/500YDA

Horizontal Cylindrical Pressure Steam Sterilizer


Saturated pressure steam is used to sterilize article, for it will liberate a great amount of heat when condensing. To make the articles (need to be sterilized) hot and wet for a certain time, most resistant bacterial spores will be killed, and then sterilization aim is reached. The method is reliable, economical, and easy to be controlled, so it has been used for more than one hundred years.

This unit is ideal equipment for clinics, scientific research institutes and other organizations to sterilize surgical instruments, fabric glassed, culture media etc.

Specifications

1.     Sterilizing course automatic sterilization controlled by computer, easy to operate.

2.     With drying function, suitable dressing drying.

3.     With over-temperature, over-pressure auto-protect device.

4.     The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.

5.     It can't be start on if the door doesn’t close well.

6.     The safety valve will be open automatically when the inner pressure over 0.24MPa, and the steam be exhausting to the water tank.

7.     The power will be cut -off automatically, cut-off the water and alarm when the machine lacking water.

Technical Data

Model technical data

WS-400YDA

   WS-500YDA

Volume of sterilizing chamber

400L, Φ700×1100mm

500L, Φ700×1300mm

working pressure

0.22MPa

working temperature

134℃

Adjustment of temperature

105℃ - 134℃

Time of sterilization

0-60 min

Time of drying

0-60 min

Heat average

≤±2℃

Power Supply with mains steam

18KW/380V 50Hz

18KW/380V 50Hz

Dimension

1430×880×1830

1800×900×1820

Transport size

1600×1050×2100

1850×1050×2100

Weight

530/420 KG

580/470 KG


Copyright © 2001-2024 ARI Medical Technology Co., Ltd. - ARI Technology Group Co., Ltd. All Rights Reserved.