Products

AR-500

Digital Mammography System

AR-500 Digital Mammography System

 

Technique data:

Main Technical Parameter                                                    

Name

Parameter

Remark

High frequency and high voltage generator 

Invert Frequency: >80kHz

nominal output current: 0~120mA

Working range: 20~35kV 10-510mAs

Over-voltage protection, over-current protection, over-power protection, overheat protection

Automatic voltage compensation

High-voltage interlocking device of filament

 

X-ray tube

 

Focus: 0.1mm/0.3mm

Target angle: 15º

Anode MaterialMolybdenum (Mo)

Port MaterialBeryllium (Be)

Anode thermal capacity: 210kJ300KHU

Inherent filtration: about 1.0mm Be

Rated tube voltage: 20~35 kV

Rated tube current: 0~120mA

 

 

Mammography film post

 

Motorized C-arm:

Electric lift post

580mm vertical movement

Rotation: +175º~-180º

Electronic compressor

        Max. pressure: 200N

Max. travel: 120mm

 

Power supply

single phase AC: 220V±10%  50Hz   6kVA

 

Specialized flat panel detector for Mammography     

 

Detector, power box, date line, capture card, etc.           

 

effective coverage of detector            

23x15cm / 29x23cm

 

Pixel matrix       

3072x1944 / 3888x3072

Limit of spatial resolution      

6.0Lp/mm

DQE value    

70%

dynamic range   

14bit digital output

pixel size        

75μm

High volta

 ge Synchronizer trigger       

BNC

Output

Camera Link or Ethernet

Working condition       

10-40

storage environment    

-10-50

Bucky

 

Housing and movement device                           

 

Size

370x275x125CM

v  ABS high-strength material          

v  High quality Brushless digital motor control      high-precision Linear guide track 

v  Perfect design with patent protection  

stepless speed regulating range  

06CM/s

Movement range         

0.52CM

Grid

 

Size

 24x30cm

v  Excellent cutting, Make grid lines and angles with excellent accuracy 

v  Newest aperture technology, have higher transmittivity 

grid ratio

3.5: 1

grid density

36L/CM

focal distance

65CM

 

 

 

DICOM

Image acquisition workstation 

CPU≥Intel Core Duo 2.60GHz                             Hardware≥250G high speed Hardware                      

memory≥2G  

display card≥512MB

19high brightness high-contrast LCD, 1280*1024 Pixel resolution

network interface  Worklist    Worklist

DICOM3.0 transmission                                   DICOM3.0

100/1000 Gigabit Ethernet                                 100/1000

 

 

 

 

 

 

Standard detector image acquisition workstation       

 