-10.4” high-resolution color touch screen
-User friendly design interface
-Tidal volume 5~1500 ml
-Up to 5 flow tubes for 3 gases
-PSU material of breathing circuit
-By-pass Tech for soda lime canister,Soda lime can be replaced during surgery
-All ventilation modes to satisfy neonate, pediatric and adult patients
-Integrated breathing circuit with built-in heater
-Highly visible ACGO
PARAMETER | SPECIFICATION |
Screen | 10.4” high-resolution color touch screen, 1024×768 pixels |
Gas supply | O2, N2O, AIR |
Flow Meter | O2: 0–10L/min
N20: 0–10L/min Air: 0–10L/min |
Selectatec bar | Dual positions |
Mechanical Flowmeter | 5 flow tubes for 3 gases |
Rapid Oxygen Flow | 25–75L/min |
Waveforms | Up to 3 waveforms, user configurable |
O2 -N2O Link System | Type: Gear wheel linkage;
Range: O2 concentration not lower than 25%. |
Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Respiratory Circuit | Working mode: all-close, semi-close, semi-open;
APL: 1-70 cmH2O |
Respiratory Bellows | Bellows for adults, tidal volume range:0–1500ml;
Bellows for children, tidal volume range:0-300ml. Anyone can be chosen for application as required by the customer |
VENTILATOR | |
Ventilation Mode | Manual, VCV, PCV, SIMV-VC, SIMV- PC, PSV, Standby, ACGO Optional modes: PRVC, HLM |
Ventilation Function | PEEP, SIGH, IP, IRV, standby |
Spirometry loop | P-V, P-F, F-V, reference Loop |
Ventilation Parameters | Tidal volume 5~1500ml
Rate 4~100bpm SIMV rate 4~40bpm I:E 4:1~1:8 Inspiratory trigger pressure OFF, PEEP-20 ~ PEEP-1cmH2O PEEP 0~70cmH2O Pressure limitation 10~100cmH2O Pplatform -20~120cmH2O Pressure control 5~70cmH2O Pressure slope 0~2 s |
Parameters for Ventilation
Monitoring |
Expired tidal volume, Inhale tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, autonomous breathing rate, I/E, peak pressure of airway, average pressure, PEEP, inspiratory trigger pressure, Inspiratory platform, oxygen concentration, battery capacity, pressure-time waveform, flow-rate-time waveform, volume-time waveform, flow-volume loop, pressure-volume loop |
Sustained High-pressure
Alarm |
Paw greater than alarm limit for continuous high airway pressure for >15s |
Auxiliary power outlets | 3 |
Wheels | Four wheels(155mm) with 2 or 4 separate brakes |
Drawers | 2 |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate.
The regulation range of evaporator concentration is 0-6vol%. Selectatec®, With interlocking function. Vaporizer types: Halothane, Enflurane, Isoflurane and Sevoflurane. |
O2 cell | Standard |
Built-in Heater | Standard |
By-pass | Standard |
ACGO | Standard |
Li-ion Battery | Standard 1 battery, 4800mAh;
Optional 2 batteries. |
Spare Cylinder yoke | Optional (O2, N2O) |
AGSS | Optional |
Gas monitor module | Optional (ETCO2, AG) |
Patient monitor | Optional |
Suction Device | Optional |
Standard: Main unit, O2 cell, Built-in Li-on battery, Disposable adult anesthesia breathing circuit, O2/N2O/Air gas pipelines, Hospital O2/N2O/Air joints | |
Options: Vaporizer (CE certified Vaporizer of the same brand), Patient monitor, ETCO2, Suction module, Pediatric/Neonate anesthesia breathing circuit, Spare cylinder yoke (O2, N2O), AGSS, Gas monitor module (ETCO2, AG), etc. |