– Output: Biphasic Truncated exponential
– Energy Sequence (child): 50J, 50J, 50J
– Energy Sequence (adult): 150J, 200J, 200J
– Charge Time: 6 sec. to 150J, 8 sec. to 200J (New battery)
– Analysis Time: 6 sec.
– The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery: Less than 10 seconds
– The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks: Less than 10 seconds
– The maximum time from initially switching power on toreadiness for discharge: Less than 18 seconds
– Visual Prompts: LED prompts; Automatic Self-test
– Controls: four buttons: On/Off, Shock, Adult/child, CPR
– Output Energy Accuracy: ±10% into any impedance from 25 to 175Q
– Output disabled when the patient impedance is outside limits: 20Ω to 200Ω.
– Part No.: CR123A-4×2
– Non-rechargeable: 2.8Ah
– Capacity: 120 discharges at 200 Joules
– Shelf Life (25℃+15C): 5 years standby (after installation)
– Dimensions: 280*209*80mm
– Weight: 1.5kg
– Operating Temperature: 0℃ to 40℃
– Operating Humidity: Relative humidity between 30% and 95% (non-condensing)
– Storage Temperature (without battery): -20℃ to 55℃
– Storage Humidity (without battery): Up to 93% (non-condensing).