Search
Close this search box.

Products

Specifications

– Output: Biphasic Truncated exponential

– Energy Sequence (child): 50J, 50J, 50J

– Energy Sequence (adult): 150J, 200J, 200J

– Charge Time: 6 sec. to 150J, 8 sec. to 200J (New battery)

– Analysis Time: 6 sec.

– The maximum time from the initiation of rhythm analysis to readiness for discharge with a

new battery: Less than 10 seconds

– The maximum time from the initiation of rhythm analysis to readiness for discharge after

6 shocks: Less than 10 seconds

– The maximum time from initially switching power on toreadiness for discharge: Less than

18 seconds

– Visual Prompts: LED prompts; Automatic Self-test

– Controls: four buttons: On/Off, Shock, Adult/child, CPR

– Output Energy Accuracy: ±10% into any impedance from 25 to 175Q

– Output disabled when the patient impedance is outside limits: 20Ω to 200Ω.

Battery

– Part No.: CR123A-4×2

– Non-rechargeable: 2.8Ah

– Capacity: 120 discharges at 200 Joules

– Shelf Life (25℃+15C): 5 years standby (after installation)

Physical

– Dimensions: 280*209*80mm

– Weight: 1.5kg

– Operating Temperature: 0℃ to 40℃

– Operating Humidity: Relative humidity between 30% and 95% (non-condensing)

– Storage Temperature (without battery): -20℃ to 55℃

– Storage Humidity (without battery): Up to 93% (non-condensing).