Type: Automatic External Defibrillator AED
Self-test: Daily, Weekly, Monthly
Mode: Adult, Child
Waveform: Biphasic truncated exponential
Energy: 200 Joule max.
Energy Sequence (programmable): 50,50,75 Joule; 150, 150, 200 Joule
Charge time (at 25℃): Less than 6 sec. to 150J; Less than 8 sec. to 200J
Voice Prompt: Extensive voice prompt
Visual Indicators: LED prompts
Control (Two buttons): ON/OFF, Shock
ECG storage: 1500 events
Data transmission: Infrared
Waveform gain: 5mm/mv、10mm/mv、20mm/mv
Waveform speed: 12.5mm/s、25mm/s、50mm/s
2 languages to choose (by end user)
5 levels of sound volume (by end user)
Adult / Child mode Button
Infra-red for data transmission (standard)
Daily, Weekly and Monthly self-test
Using the AED while in bag (don’t need to open the bag while using the machine)
Extra pocket of carry bag for storing first aid products
IPX55 water and dust resistant
Three-step defibrillation process defibrillator
Selectable Energy for different age patients: 50, 50, 75 Joule; 150, 150, 200 Joule
Extensive voice and visual prompts for the operator
Changeable language manually
Biphasic energy output
Lock-out protection to prevent inadvertent defibrillation
Device has an internal memory (60 hours)
ECG signals and events can be transferred to the computer
Provided with carrying case, battery, pads, software for data transmission
Long-life disposable LiMnO2 type battery (battery life of at least 5 years)
It’s used to assist the supervisory system of Automatic External Defibrillator
The interface display status information recorded in the running of the AED, including type, event time, description, waves.
Event type includes on event, self-test event, analyzing event, charging event, discharging event, CPR starting event, CPR stopping event. The event time displays when the event happens.
Description shows the parameter information. “OHM” represents the detected impedance of the patient. Power parameter represents the voltage of the finished charging. “Need to shock” represents that shock advised after analyzing. “No need to shock” represents that no shock advised after analyzing. Waveform represents the link of the ECG waves.
Click the “print” in ECG waveform display interface, then enter the setting interface of print.
Power: 12V, 3000mAh
Type: Non-rechargeable Li-MnO2 Cell
Environmental Operating Temperature: 0℃ to 40℃
Operating Humidity: Relative humidity between 30% and 95% (non-condensing)
Storage Temperature (without battery): -20℃ to 55℃
Storage Humidity (without battery): Up to 93% (non-condensing)
Weight (with battery): 1.9Kg (4.2 pounds)
AED Pad 1Pcs
Battery 1Pcs
AED Case 1Pcs
Cleaning Cloth 1Pcs
CD Software for ECG Data transfer
AED Cabinet
USB Adapter for ECG Data transfer
Copyright © 2001-2025 ARI Medical Technology Co., Ltd. - ARI Group. All Rights Reserved.
Copyright © 2001-2025 ARI Medical Technology Co., Ltd. - ARI Group. All Rights Reserved.