PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of Application Adult, child
Gas Source O2: 0.28–0.6MP
Flow Meter O2: 0.05–1.0L/min; 1.1–10L/min
Flow Rate of Rapid Supply 25–75L/min
Low Oxygen Pressure Alarm It will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.

Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.

Respiratory Circuit Working mode: all-close, semi-close, semi-open

APL: ≤12.5kPa.

Respiratory Bellows Bellows for adults, bellows for children, tidal volume range:0-1500ml
VENTILATOR
Display Mode High-definition 5.7” LCD screen display
Ventilation Mode IPPV, SIPPV, SIMV, MANUAL
Ventilation Function PEEP, SIGH, IRV
Ventilation Parameters Tidal volume           50~1500ml

Rate                  2~99bpm

SIMV rate                   2~20bpm

I:E                  2:1~1:8

SIGH          1.5 times the inspiratory time/100bpm

Inspiratory trigger pressure   -10~20cmH2O

PEEP                 2~20cmH2O

Pressure limitation      10~60cmH2O

Parameters for Ventilation

Monitoring

Tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway-pressure-time waveform, flow rate-time waveform, PEEP, inspiratory trigger pressure, (average pressure, oxygen concentration, battery capacity).
SECURITY ALARM SYSTEM
Oxygen Concentration Alarm Upper limit setting range   (21%~99%)

Low limit setting range     (18%~80%)

Airway Pressure Alarm Upper limit setting range   55~60cmH2O

Low limit setting range     0~50cmH2O

Minute Ventilation Volume Alarm Upper limit setting range   1~20L/min

Low limit setting range     0~19L/min

Tidal volume Upper limit setting range   100~1500ml

Low limit setting range     0~1400ml

Sustained High-pressure

Alarm

It will give alarm when stress have consistently been higher than

25cmH2O

(Suffocation Alarm, Power alarm, Intubation off alarm, battery voltage alarm)
Power AC 220V, 50Hz, (UPS, Storage battery)