PARAMETER | SPECIFICATION |
Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of Application | Adult, child |
Gas Source | O2: 0.28–0.6Mpa;
N2O: 0.28–0.6MPa. |
Rapid Oxygen Flow | O2: 0.05–1.0L/min; 1.1–10L/min
N20: 0.05–1.0L/min; 1.1–10L/min |
O2, N2O Linkage and N2O
Stopper |
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure < 0.2KPa, the flow of nitrous oxide would be cut off. |
Flow Rate of Rapid Oxygen Supply | 25–75L/min |
Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
Respiratory Circuit | Working mode: all-close, semi-close, semi-open;
APL: ≤12.5kPa. |
Respiratory Bellows | Bellows for adults, bellows for children, tidal volume range: 0-1500 ml. |
VENTILATOR | |
Display Mode | High-definition 5.7” LCD screen display |
Ventilation Mode | IPPV, SIPPV, SIMV, MANUAL |
Ventilation Function | PEEP, SIGH, IRV |
Ventilation Parameters | Tidal volume 50~1500ml
Rate 2~99bpm SIMV rate 2~20bpm I:E 2:1~1:8 Inspiratory trigger pressure -10~20cmH2O PEEP 2~20cmH2O Pressure range 10~60cmH2O SIGH 1.5 times the inspiratory time/100bpm |
Parameters for Ventilation
Monitoring |
Tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure-time waveform, flow rate–time waveform, PEEP, inspiratory trigger pressure, average pressure, oxygen concentration, battery capacity. |
SECURITY ALARM SYSTEM | |
Oxygen Concentration Alarm | Upper limit setting range (21%–99%);
Low limit setting range (18%–80%) |
Airway Pressure Alarm | Upper limit setting range 5–60cmH2O;
Low limit setting range 0–50cmH2O. |
Minute Ventilation Volume Alarm | Upper limit setting range 1–20L/min;
Low limit setting range 0–19L/min. |
Tidal Volume | Upper limit setting range 100–1500ml
Low limit setting range 0–1400ml |
Sustained High-pressure
Alarm |
It will give alarm when stress have consistently been higher than
25 cmH2O |
Suffocation Alarm | Power alarm, Intubation off alarm, Battery voltage alarm |
Power | AC 220V, 50Hz, Storage battery, UPS |
Mechanical Arms | |
(End-expiratory CO2 monitoring) | |
(Specific interface of waste anesthesia gas) | |
*Note: Items within brackets are optional. |
Copyright © 2001-2025 ARI Medical Technology Co., Ltd. - ARI Group. All Rights Reserved.