PARAMETER | SPECIFICATION |
Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of Application | Adult, child |
Gas Source | O2, N2O, Air: 0.28–0.6MPa. |
Flow Meter | O2: 0.05–1.0L/min; 1.1–10L/min
N20: 0.05–1.0L/min; 1.1–10L/min Air: 0.05–10L/min |
O2, N2O Linkage and N2O Stopper | When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure < 0.2MPa, the flow of nitrous oxide would be cut off. |
Rapid Oxygen Flow | 25–75L/min |
Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
Respiratory Circuit | Working mode: all-close, semi-close, semi-open;
APL: ≤12.5kPa |
Respiratory Bellows | Bellows for adults, tidal volume range:0–1500ml;
Bellows for children, tidal volume range:0-300ml. Anyone can be chosen for application as required by the customer |
VENTILATOR | |
Display Mode | High-definition 10.4” TFT color LCD screen display |
Ventilation Mode | VCV, PCV, SIMV, MANUAL |
Ventilation Function | PEEP, SIGH, IP, IRV, standby |
Ventilation Parameters | Tidal volume 20~1600ml
Rate 2~99bpm SIMV rate 2~20bpm I:E 4:1~1:8 Inspiratory trigger pressure -10~20cmH2O PEEP 2~20cmH2O Pressure limitation 5~60cmH2O SIGH 1.5 times the inspiratory time, 60~120 adjustable Inspiratory platform 0~50% Pressure control 3~50cmH2O Pressure slope 1~10 gear |
Parameters for Ventilation
Monitoring |
Expired tidal volume, Inhale tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, autonomous breathing rate, I/E, peak pressure of airway, average pressure, PEEP, inspiratory trigger pressure, Inspiratory platform, oxygen concentration, battery capacity, pressure-time waveform, flow-rate-time waveform, flow-volume loop, pressure-volume loop |
SECURITY ALARM SYSTEM | |
Oxygen Concentration Alarm | Upper limit setting range 21%–99%;
Low limit setting range 18%–80% |
Airway Pressure Alarm | Upper limit setting range 5–60cmH2O;
Low limit setting range 0–50cmH2O. |
Minute Ventilation Volume Alarm | Upper limit setting range 1–20L/min;
Low limit setting range 0–19L/min. |
Respiratory Rate | Upper limit setting range 10–99bPm
Low limit setting range 0–50bPm |
Tidal Voulme | Upper limit setting range 10–1500ml
Low limit setting range 0–1400ml |
Sustained High-pressure
Alarm |
It will give alarm when stress have consistently been higher than
25 cmH2O |
Suffocation Alarm | Power alarm, Battery voltage alarm, Intubation off alarm |
Power | AC 220V, 50Hz, UPS, Storage battery |
Manipulators | |
(End-expiratory CO2 monitoring) | |
(Specific interface of waste anesthesia gas) | |
*Note: Items within brackets are optional. |
12-inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR