Search
Close this search box.

Products

Features

-Throughput 60T/H
-8-inch touch screen
-1 press for analysis with 20 parameters + 3 histograms
-Sample volume:10μL
-2 reagents +1 probe cleaser only
-3 counting modes
-10,000 sample results
-Support LIS and external printer

Technical Specifications

PARAMETER SPECIFICATION
Principle Impedance for cell counting and cyanide-free method for HGB
Test flow 1 click to finish test automatically,10μL blood is needed
Parameters 3 part differentiation of WBC,20 parameters +3 histograms
Report parameters WBC,LYM%,MID%,NEUT%,LYM#,MID#,NEUT#,RBC,HGB,HCT,MCV, MCH, MCHC,RDW-SD,RDW-CV, PLT, MPV, PDW, PCT, P-LCR
Histograms WBC, RBC, PLT histogram
Throughput 60 test/Hour
Languages English, Spanish,Portuguese, etc.
Calibration Auto mode and manual mode
Quality control 3 level QC, LJ graph
Sample volume

 

Venous mode: 10μL venous blood
Capillary mode: 10μL capillary blood
Prediluted mode: 20μL capillary blood
Display 8-inch color LCD with touch screen
Resolution: 800×600
Storage 100,000 sample results with histograms
Reagent 2 reagents:Lyse(500mL) and Diluent(20L)
Printout Inner thermal printer, support external printer
Input/Output Keyboard,mouse,RS232 serial port
Power AC 100-240V, 50/60±1Hz
Dimension 410mm×435mm×472mm
Reference range type Newborn, children,male,female and general

Performance

=Parameters Precision (CV)
WBC ≤ 2.0%    (7.0 -15.0×109)/L
RBC ≤ 1.9%    (3.5 – 6.0×1012)/L
HGB ≤ 1.9%    (110.0 – 180.0g)/L
MCV ≤ 0.4%    (80.0 – 110.0) fL
PLT ≤ 4.0%    (100.0 – 500.0×109)/L
=Parameters Measurement Range
WBC (0-99.9×109)/L
RBC (0-9.99×1012)/L
HGB (0-300.0)g/L
PLT (0-999×109)/L