PARAMETER | SPECIFICATION |
Sample/Reagent position | 40pcs sample position |
Reagent refrigerated function | |
40pcs reagent position | |
Reacting system | Reaction cuvette:60 reusable plastic units |
Optical length of cuvette:6mm | |
Max reaction time:10 minutes | |
Reaction temperature:37±0.1°C | |
Reaction cuvettes | Reaction Cuvette: 60 reusable plastic cuvettes |
Automatic blank testing for cuvette | |
Colorimetry system in the pattern of light splitting behind cuvette | |
Dismountable,easily replacement | |
Optical System | Light Source: halogen-tungsten lamp |
Wavelength: 340-690nm,9 wavelengths,precision ±1.5nm | |
Absorbance range:0~5.0 Abs | |
Resolution:0.0001 Abs | |
Rear spectrophotometry | |
Optimum calibration
cuvette |
Automatic cuvette blank testing |
Automatic wash selected cuvettes | |
Calibration | Calibration: linear/nonlinear multi-points calibration |
Re-test: retest the sample automatically when the result is out of the linearity range or the sample is not sufficient | |
Sample Reagent/Mixing
probe |
Automatic liquid level detect |
Collision protection function | |
Probe internal/external washing | |
Independent mixing probe | |
7-step auto cleaning
system |
Automatic 7-step washing |
Hi-tech washing station to ensure accurate results | |
With low water consumption:1-2L/H | |
Intellective software | Automatic washing cuvette when startup&shut down |
Automatic detect reagent volume during every startup | |
Reagent plate temperature detecting when startup | |
Dynamic and real-time
display of running status |
Real-time online monitoring running status of sample tray, reagent tray and cuvette |
Real-time online monitoring reagents residual volume | |
Real-time online monitoring waste volume | |
Automatic dilute and
re-test |
Automatic dilute and re-test when absorbance range is out of range |
Automatic dilute and re-test when linear limit is out of range | |
Automatic dilute and re-test when linear range is out of range | |
Freely set the auto dilute ratio and dilute water position | |
Size | 54.8cm×42.2cmX38.3cm |
Net weight | ≤23kg |
Optimal substitution of
semi-auto biochemistry analyzer
|
Minimal labor required |
Independent mixing probe | |
Utility of reagents decreased:300ul/test | |
Efficiency speed:200 tests/hour | |
User friendly: Smart and easy-operating software | |
High Accuracy: Minimal operating/random errors | |
High reputation from clinics | |
System functions
|
Analysis Methods: end-point, kinetic, two-point, double-reagent, double-wavelength, multi-standard etc, open to various reagents |
Sample/Reagents Handling | Sample volume:1.6-50μl,0.1μl step |
Sample Dilution: automatic/Manual pre-dilution, with
dilution ratio up to 1:100 |
|
Reagent Position: 40pcs | |
Reagent Volume:10-500μl,0.5μl step | |
Working condition | Power Supply:100-240V,50/60Hz,300VA |
Temperature:10-35℃ | |
Humidity:≤90% no dew | |
Operation system | Windows 10 or Windows 7 |
Input and Output | Input: RS-232 interface/computer |
Output: multi-format printout |
Copyright © 2001-2025 ARI Medical Technology Co., Ltd. - ARI Group. All Rights Reserved.