PARAMETER SPECIFICATION
Gas Supply
Gas Supply Medical-grade Oxygen
Pipe Connector NIST or DISS
Gas Supply Pressure 280~600kPa
Display
Display Size 12.1 inch-TFT Touch Screen
Display Resolution (H)*(V) 1280*800 pixeles
Brightness Adjustable
Communication Interface
Communication Interface Standard: HDMI, RS232, USB, Nurse Call

Optional: CO2 Module, SPO2 Module

Power & Battery Backup
AC Power Supply Input Voltage 100~240V
Input Frequency 50/60Hz
Input Current 2.5A Max
Fuse T3.15AH/250V
Number of Batteries One or Two (Two Batteries is Optional)
Battery Type Built-in Lithium-ion Battery, 14.4VDC, 6900mAh
Battery Running Time 3 hours (1pc Battery), 6 hours (2pcs Battery)
Value Added Functions
100% O2

Suction

Nebulization*

Manual Breath*

Inspiratory Hold*

Expiratory Hold*

PEEPi*

P0.1*

NIF*

PV Aid

Dynamic Lung

NOTE: * Only available on AR-V200

Note: Flow Sensor is Disposable

Ventilation Specifications

PARAMETER SPECIFICATION
Patient Type Adult, Pediatric, Neonatal
Invasive Ventilation Mode – VCV (Volume Control Ventilation)

– PCV (Pressure Control Ventilation)

– VSIMV (Volume Synchronized Intermittent Mandatory Ventilation)

– PSIMV (Pressure Synchronized lntermittent Mandatory Ventilation)

– CPAP/PSV (Continuous Positive Airway Pressure/Pressure Support Ventilation)

– PRVC (Pressure Regulated Volume Control)

– V+SIMV (PRVC+SIMV)

– BPAP* (Bilevel Positive Airway Pressure)

– APRV* (Airway Pressure Release Ventilation)

– Apnea Ventilation

– AMV

Non-invasive Ventilation Mode – PCV (Pressure Control Ventilation)

– CPAP/PSV (Continuous Positive Airway Pressure/Pressure Support Ventilation)

– BPAP* (Bilevel Positive Airway Pressure)

– APRV* (Airway Pressure Release Ventilation)

– PSIMV (Pressure Synchronized lntermittent Mandatory Ventilation)

NOTE: * Only available on AR-V200
Apnea Ventilation
VT-apnea Adult: 100~2000mL

Pediatric: 20~300mL

Neonatal: 2~300mL

△P-apnea 5~60cmH2O
f-apnea 1~80/min
Apnea T-insp. 0.2~10s
Sigh
Sigh Switch On, Off
Interval 20s~180min
Cycles Sigh 1~20
△int. PEEP 0~45 cmH2O, Off
O2 Therapy O2 Flow
Controlled Parameters 21~100% (Increments of 1%) 4~60L/min
Controlled Accuracy ±(3vol.%+1% of Setting) ±(2L/min+10% of Setting) (BTPS)

Controlled Parameters

SPECIFICATION ADULT, PEDIATRIC NEONATAL
O2% 21~100% 21~100%
VT (Tidal Volume) 100~2000mL (Adult)

20~300mL (Pediatric)

2~300mL
f 1~80/min 1~150/min
f-SIMV 1~80/min 1~150/min
I:E 1:10~4:1 1:10~4:1
T-insp. 0.2~10s 0.10~10s
T-slope 0~2s 0~0.6s
T-high 0.2~30s 0.2~30s
T-low 0.2~30s 0.2~30s
T-pause 5%~60%, Off 5%~60%, Off
△P-insp. 5~80cmH2O 3~60cmH2O
△P-supp. 0~80cmH2O 0~45cmH2O
P-high 0~80cmH2O 0~45cmH2O
P-low 0~45cmH2O 0~25cmH2O
PEEP 1~45cmH2O, Off 1~25cmH2O, Off
Flow Trigger 0.5~15L/min 0.1~5L/min
Pressure Trigger -10~(-0.5) cmH2O /
Exp% 10~85%, Auto 10~85%, Auto

Synchronized Tube Resistance Compliance

PARAMETER SPECIFICATION
Tube Type ET Tube, Trach Tube, Disable STRC
Tube I.D. Adult: 5.0~12.0mm

Pediatric: 2.5~8.0mm

Compensate 0~100%
Expiration Compensation Switch On, Off

Monitored Parameters

Numeric
Paw Vte Ri
P-peak VTi C-dyn.
P-plat. VTe-spn. C-stat.
P-mean VTe/IBW RC-exp.
PEEP FiO2 WOB
MV f-total RSBI
MV-leak I:E Continuous Flow (O2 Therapy)
MV-spn. Re
Real Time Graphics
Pressure-time Waveform Pressure-volume Loop
Flow-time Waveform Flow-time Loop
Volume-time Waveform Pressure-flow Loop
Trend & Record  
Type Tabular, Graphic
Length 72 hours
Content Monitor Parameters
Max Log 20000
Screen Capture 100

Alarm Settings

SPECIFICATION TYPE ADULT PEDIATRIC NEONATAL
Tidal Volume High

Low

110~4000mL, Off

50~4000mL, Off

25~600mL, Off

10~600mL, Off

4~600mL, Off

1~590mL, Off

Minute Volume High

Low

0.2~100.0L/min

0.1~50.0L/min

0.2~60.0L/min

0.1~30.0L/min

0.12~50.0L/min

0.02~30.0L/min

Airway Pressure High

Low

10~65cmH2O

10~63cmH2O, Off

10~65cmH2O

10~63cmH2O, Off

10~65cmH2O

10~63cmH2O, Off

Frequency High

Low

1~90/min, Off

1~88/min, Off

1~90/min, Off

1~88/min, Off

1~160/min, Off

1~158/min, Off

O2 Concentration High

Low

20~100%

18~98%

20~100%

18~98%

20~100%

18~98%

Apnea Alarm Time Low 5~60s 5~60s 5~60s

Physical Specifications

PARAMETER SPECIFICATION
Dimensions (L*W*H) 336mm*330mm*345mm

664mm*600mm*1310mm (with Trolley)

Weight Approximately 9.9kg (Main Unit), Approximately 29.5kg (with Trolley)
Environmental Conditions
OPERATING STORAGE AND TRANSPORT
Temperature 5~40℃ -20~60℃, O2 Sensor: -20~50℃
Relative Humidity 10~95% 10~95%
Barometric Pressure 62~106kPa 50~106kPa