PARAMETER | SPECIFICATION |
Product name | Ethylene Oxide Sterilizer |
Product form | Vertical |
Heating method | Electric heating |
Power
Volume Chamber size |
H80: 2.6KVA / 67L (Liner size: 585 x 275 x 422mm)
H120: 3.0KVA / 103L (Liner size:585 x 375 x 472mm) H220: 3.3KVA / 196L (Liner size:705 x 505 x 552mm) H330: 3.6KVA / 294L (Liner size:760 x 575 x 682mm) |
Power | AC220V, 50Hz |
Temperature display | 3.5 inch LCD display |
Material of Sterilization Room | 304 anti-corrosion stainless steel |
Outer box material | Carbon steel spray plastic |
Temperature during sterilization | 50℃±3℃ |
Working pressure | -60kpa |
Highest anti-stress value | -140kpa |
Humidity during sterilization | 45%-80% |
Selected gas | Pure ethylene oxide gas |
Sterilization time | 0-99 hours adjustable |
Air Replacement | Automatic ventilation, 5 cycles of ventilation, every 15 minutes |
Full-automatic control system | One-click start to finish, the whole process is completed
automatically |
Sterilization record
|
The system can store 500 pots of sterilization records |
Display status
|
Display operating status, temperature, humidity, pressure, sterilization time, residual gas treatment time, times, steam generator temperature. |
Humidity Sensor | The built-in humidity sensor automatically detects the humidity in the cavity. |
Pressure Sensor | The built-in pressure sensor automatically detects the pressure in the cavity |
Temperature Sensor | The built-in temperature sensor automatically detects the temperature in the cavity, the accurate range is 50℃±3℃. |
Vacuum system | Use oil-free vacuum pump. |
Seal door | 304 stainless steel aluminum material, thickness ≥ 2mm, pressure resistance 200Kpa. |
Humidification System | Using pulse steam humidification |
Residual gas treatment device | The filtration rate of the filter to the particle diameter ≥0.3μm is
not less than 99.5% |
Vacuum rate | Under no-load conditions, the time for the sterilizer to be vacuumed from normal pressure to -60 kPa ≤ 30 min |
Automatic dosing | Puncture technology built into the sterilization chamber. |
Heating system | The explosion-proof electric heating device is used in the sterilization chamber |
Multiple protection functions | With over-voltage and over-temperature protection functions, it is
more safe and reliable |
Alarm system | Have high temperature alarm system, ultra low pressure alarm system |