ARI Vetcare delivery high precious veterinary anesthesia equipment system. It is suitable for animal hospital, pet clinic and animal laboratory. This technical index of animal anesthesia machine can meet the needs of general anesthesia and medical research on mice, dogs, cats, rabbits, monkeys, pigs, sheep, and other animals in animal laboratory.

Technical Specifications

PARAMETER SPECIFICATION
Application 0.5-100kg animal
Ventilation mode Closed, half-closed, half-open
Tidal volume 10-2000ml
Vaporizer Choose one from Isoflurane, Sevoflurane, Halothane
Oxygen source pressure 0.25~0.65Mpa
Airway pressure scope 030~60CMH2O
Flow Control 0-4L/min
Oxygen Flush 35L/min~75L/min
Pop adjust scope Directs the waste anesthetic gas from the machine to a scavenging system. In the completely open position, the pop-off valve will release pressure at 2 cm H2O, while maintain a constant passive volume in the breathing bags.
Inspiration impedance ≦0.6Kpa
Expiration impedance ≦0.6Kpa
Absorption tank capacity 700ml
Monitoring Airway pressure, Gas source pressure monitoring
I:E As doctor need
BPM As doctor need

Configuration

Standard: Main unit, one anesthesia vaporizer, one set veterinary breathing circuit, one mobile trolley, breathing circuit, gas source hose, exhaust gas absorption system, corrugated connect pipeline, simulated lung.

Optional: Anesthesia ventilator (V-60V, V-80V), non-breathing circuit, tracheal intubation, anesthesia laryngoscope, soda lime, active carbon.

Anesthesia Ventilators (Optional)

Option 1: V-60V Anesthesia Ventilator

Pneumatically driven and electrically controlled

Display: LCD digital display

Animal Bellow: Big animal: 300-1500ML; Small animal: 0-300ML

Monitoring: Tidal volume, breathing rate, I:E ratio, peak airway pressure

TV: 0-300, 300-1600ml

Breathing rate: 2~150bpm

I:E ratio: Adjustable 3:1,2:1,1.5:1,1:1,1:1.5,1:2,1:3

Upper airway pressure:2~6kpa

Lower airway pressure:0.6~2kpa

Build in battery: work time more than 1 hour in case of power failure

Alarm: Battery low, Power failure, Airway pressure high/low

Alarm Silence: Manuel/Auto Switch

PIP scope: 5-60kpa

 

Option 2: V-80V Anesthesia Ventilator

Display: 9-inch TFT touch screen

Drive gas:

Gas type: O2 or Air

Inlet pressure: 280-600 kPa

Max flow: ≤120 L/min

Operating mode: VCV, PCV, SPONT, Manual, Demo

Waveform: Pressure, Flow-rate, Volume