M-9000E

Patient Monitor

Feature

1.    12.1-inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR

2.    13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;

3.    Efficient resistance to interference of defibrillator and electrosurgical cautery;

4.    SPO2 can testing for 0.1% Weak;

5.    RA-LL impedance Respiration;

6.    Trend Coexist Display;

7.    OxyCRG DyNamic View Display;

8.    Bed to Bed view Display;

9.    Networking capacity and nurse calling system;

10.  Built-in rechargeable battery 4400mAh;

11.  Large volume of tabular and graphic trends information storage and easy to recall;

12.  Anti-ESU, anti- defibrillator;

13.  Capture dynamic waveforms.

Technical Specification

ECG

Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)

Lead selection: I, II, III, avR, avL, avF, V,

Waveform: 2 ch

Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto

Heart Rate and Alarm

Range

Adult: 15 ~ 300 bpm

Neo/Ped: 15 ~ 350 bpm

Accuracy: ± 1% or ± 1bpm, which great

Resolution: 1 bpm

ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV

Respiration

Method: Impedance between R-F (RA-LL)

Bandwidth: 0.3 ~ 2.5 Hz

Resp. Measuring Range:

-Adult: 0 ~ 120 rpm

-Neo/Ped: 0 ~ 150 rpm

Resp. Alarm Range:

 -High alarm range limit: 8~120 rpm

 -Low alarm range limit: 6~120 rpm 

Resolution: 1 rpm

Accuracy: ±2 rpm

NIBP

Method: Oscillometric

Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)

Measuring Period in STAT Mode: 5 Min

Alarm Type: SYS, DIA, MEAN

Resolution Pressure: 1mmHg

Overpressure Protection

-Adult Mode: 297±3 mmHg

-Pediatric Mode: 240±3 mmHg

-Neonatal Mode: 147±3 mmHg

SpO2

Measuring Range: 0 ~ 100 %

Resolution: 1 %

Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified

Pulse Rate

Measuring and Alarm Range: 20~300bpm

Resolution: 1bpm

Accuracy: ±2bpm

Temperature

Channel: 2

Measuring and Alarm Range: 0 ~ 50 °C

Resolution: 0.1°C

Accuracy: ±0.1°C

IBP(optional)

Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2

Measuring and alarm range

-ART: 0 ~ 300 mmHg

-PA: -6 ~ 120 mmHg

-CVP/RAP/LAP/ICP: -10 ~ 40 mmHg

-P1/P2: -10 ~ 300 mmHg

Resolution: 1 mmHg

Accuracy: ±2% or ±1mmHg, which great

Standard Packing List

5-lead ECG cable (including RESP)

Adult finger SpO2 sensor   

Adult NIBP cuff   

NIBP extension tube 

TEMP sensor 

AC power adaptor

ECG electrodes

Operation instruction

Optional Function

Internal Recorder, IBP, EtCO2, Central Monitoring System, Touch Screen, Neonate SPO2