Veterinary Syringe Pump

Features

1.      Manual or Automatic Bolus/Purge/KVO function.

2.      Vertical & Horizontal mounting clamp.

3.      DPS dynamic pressure display.

4.      Anti-bolus system.

5.      Large LCD display with adjustable brightness.

6.      2000 events stored can be checked/downloaded.

Technical Specifications

PARAMETER

SPECIFICATION

Applicable syringes

5ml,10ml, 20ml, 30ml, 50ml, 60ml of any standard

Accuracy

+/- 2% with correct calibration

Rate mode

0.1ml/h-2000.0ml/h (0.1ml/h or 1ml/h step)

Time Mode

SOL. VOL: 0.1-9999.9ml (0.1ml step)
TIME: 00:01-99:59min (1min step)

Dose Mode

Volume: 0.1-999.9ml (0.1ml step)
DRUG MASS: 0.1-999.9mg (0.1mg step)
DOSE: 0.001-9999 (0.001 step)
WEIGHT: 0.1-300kg (0.1kg step)

Total Volume Range

0.1-9999.9ml (0.1ml step)

Drug library Mode

(Optional)

Sequential Mode

(Optional)

Purge Rate

Programmable and adjustable

KVO Rate

0.0~5.0ml/h, Programmable and adjustable

DPS Occlusion Pressure

LOW: 60(±15) kPa, MEDIUM: 90(±15) kPa

HIGH: 120(±15) kPa (Adjustable)

Alarms

Loose, Occlusion, Empty, Near Empty, Rate Abnormal, Low Battery,

No Battery, Infusion Complete, No Action, Syringe Plunger Disengagement

Alarm Volume

Three levels: LOW, MEDIUM or HIGH

Bolus Rate

0.1-2000ml/h, Programmable and adjustable,

Automatic/Manual bolus available

Bolus Volume

Range: 1.0~50.0ml, (0.1ml step)

Interface

RS-232

Brightness

1~10 levels can be selected

History Records

2000 events stored can be checked and downloaded

Power Source

AC power supply: AC 100 to 240V (±10%), 50/60Hz;

DC Power supply: 12~15V, power>15W;

Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh

12 hours of continuous operation

Protection Class

Class I, Type CF, IPX4

Dimension/N.W.

240mm*150mm*117mm, 2.5kg

Packing Size/G.W.

200mm*260mm*160mm, 3.5kg