Products
-
- Operating Theatre Equipment
- OB/GYN Equipment
- Diagnostic Equipment
- Sterilizer (Autoclave)
- X-ray Series
- Laboratory Equipment
- Biological Microscope
- Spectrophotometer
- High Speed Centrifuge
- Low Speed Centrifuge
- High Speed Freeze Centrifuge
- Low Speed Freeze Centrifuge
- Special Centrifuge
- Thermostatic Bath
- Laboratory Shaker
- Laboratory Mixer
- Water Distiller
- Magnetic and Electric Stirrer
- Laboratory Incubator
- Dry Oven
- PH Meter
- Biological Safety Cabinet
- Clean Workbench
- Water Purifier
- Ultrasonic Cleaner
- Pipette
- Microtome
- Medical Hand Sink
- Medical Freezer
- Blood Bank Refrigerator
- Medical Pharmacy Refrigerator
- Analysis Balance
- Medical Destroyer
- In-Vitro Diagnostics
- Hemodialysis Equipment
- Hospital Furniture
- Walking Aids
- First-Aid Products
- Medical Endoscope
- Medical Human Model
- Ophthalmic Equipment
- Dental Equipment
- ENT Equipment
- Home Care Equipment
- Veterinary Equipment
- Medical Consumables
- Personal Protective Equipment
Overview
The ARI-L series patient monitor is designed to meet daily clinical needs and seamlessly integrate into the hospital workflow. A variety of monitor models have more and more accurate practical clinical application values. In acute care, patient monitors must be reliable, easy to use, have advanced parameters, and allow access to data when needed. When transporting patients, the equipment should be easy to carry. The ARI-L series of patient monitors are lightweight, powerful, and intuitive in user interface. They are the best choice for acute, in- and out-of-hospital transfers.
Features
-The monitor adopts Ergonomic design and a compact structure to meet the needs of various parameter detections. It is suitable for adults, children, and newborns.
-Standard parameters: SpO2, 3/5 leads ECG, PR, HR, RESP, NIBP, 2*TEMP.
-Optional parameters: EtCO2, 12 leads ECG, 2*IBP.
-10/12/15 inch color LCD screen with optional touch screen and rotator knob.
-Capable to visualize 6 or more waveform. Support color coding for different waveform.
-28 ST segment analysis with ST trend with continuous display of ST value on screen.
-Multiple screen layout to view big font size in numeric and waveform.
-Trending facilities for up to 360 hours of both graphical & numerical along with event review facility for all parameters and full waveform disclosure facility.
-Different visual alarm with color coding and audible alarms for various parameters with three levels of volume adjustment.
-Support HL7 protocol to access the central monitoring system, low-power design, extending battery life Configure, fanless design to ensure a quieter and cleaner environment Arrhythmia detection and analysis in various leads.
Product Performance
-Standard application: 5-leads ECG, NIBP, SpO2, RESP, TEMP, PR/HR
-Optional application: Touch screen, IBP, EtCO2 (mainstream/sidestream), 12-lead ECG, recorder, CMS
-Large font, dynamic short trend display interface, calculation of hemodynamics and cardiac output.
Specifications
Display
Size: ARI-L10: 10 inches Color TFT LCD
ARI-L12: 12 inches Color TFT LCD
ARI-L15: 15 inches Color TFT LCD
Resolution: ARI-L10/L12: 1024*768 pixels
ARI-L15: 1366*768 pixels
ECG
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
ECG Range: 0~100%
ECG Resolution: 2.36uV/LSB
HR Range: 15~300bpm(adult), 15~250bpm(pediatric/neonate)
HR Accuracy: ±1 bpm or ±1%, whichever is greater.
RR Range: 0~120bpm
RR Accuracy: 15~120rpm: ±2rpm or ±2% <15rpm: Undefined
Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s, less than ±5% error
Display sensitivity: 2.5 mm/mV (X0.25), 5 mm/mV (X0.5), 10 mm/mV (X1), 20 mm/mV (X2), 40
mm/mV (X4), Auto, less than ±5% error
SPO2
SPO2 Measurement Range: 0~100%
SPO2 Accuracy: 70 to 100%: ±2% (adult/pediatric, without motion)
70 to 100%: ±3% (neonate, without motion)
0 to 69%: Not specified.
PI Range: 0~20%
PR Range: 25-250bpm
PR Accuracy: ±2bpm or ±2% (whichever is greater)
TEMP
Technique: Thermal resistance
Measurement Range: 0~50.0℃
Accuracy: ±0.2°C
Resolution: 0.1°C
Refresh Rate: 1s
Alarm limit range: Temp High: (low limit +0.1℃) ~ 50.0℃
Temp Low: 0℃ ~ (high limit -0.1℃)
NIBP
Measure method: Automatic oscillation method
Pressure Range: 0~300mmHg (0~40kPa)
Pressure Accuracy: ±2mmHg or ±1% of reading (whichever is greater)
Resolution: 1mmHg or 0.1kPa
SYS Range: Adult: 40-270mmHg
Pediatric: 40-200mmHg
Neonate: 40-130mmHg
DIA Range: Adult: 10-210mmHg
Pediatric: 10-162mmHg
Neonate: 10-90mmHg
Mean Range: Adult: 20-230mmHg
Pediatric: 20-170mmHg
Neonate: 20-100mmHg
Accuracy: mean deviation<±5mmHg, standard deviation<8mmHg
Work pattern: manual, automatic, continuous measurement
Automatic mode measurement interval: 1min/ 2min/ 3min/ 4min/ 5min/ 10min/ 15min/ 30min/ 60min/ 90min/ 2h/ 3h
Continuous mode measurement time: 5 min
Maximum single measurement time: 180s
Respiration
Technique: Trans-thoracic impedance
Lead: Options are lead I and II
Respiration excitation waveform: <300µA RMS, 62.8kHz (±0%)
Respiration impedance range: 0.3~5Ω
Baseline impedance range: 200~2500Ω (using an ECG cable with 1kΩ resistance)
Bandwidth: 0.2~2Hz (-3 dB)
Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s
PR
Measurement range: 25bpm~250bpm
Resolution: 1bpm
Alarm limit: PR High: (PR Low limit +1bpm) ~ 300bpm
PR Low: 15bpm ~ (PR High limit -1bpm)
HR
Measurement range: 15~300bpm (adult)
15~350bpm (pediatric/neonate)
Resolution: 1bpm
Accuracy: ±1 bpm or ±1%, whichever is greater
Sensitivity: 0.2mV
Data Storage
Trends: A minimum of 120 hours trend data with the resolution no less than 1 minute
Events: 200 events
NIBP measurements: 200 events
Interpretation of resting 12-lead ECG results: 20 sets
IBP (Optional)
Number of channels: 2
Pressure Range: -50~350mmHg (-6.7~40.0kPa)
Accuracy: ±2% or ±1 mmHg, whichever is greater (excluding sensor error)
Resolution: 1 mmHg (0.1kPa)
Refresh Rate: 1s
Sensitivity: 5μV/V/mmHg
Zero adjustment range: ±200 mmHg
Impedance range: 300~3000Ω
Volume displacement: <0.04mm³/100mmHg
EtCO2 (Optional)
Measurement mode: Sidestream, Mainstream
Technique: Infrared absorption
Zero RR delay: 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
CO2 Range: 0~20.0 vol%
CO2 Accuracy: 0~12 vol%: ± (0.2 vol%+2% of readings)
12~20 vol%: ± (0.2 vol%+6% of readings)
AwRR Measurement Range: 0~150rpm
Resolution: 1 rpm
AwRR Accuracy: Mainstream: 0~150rpm, ±1rpm
Sidestream: 0~69rpm, ±1rpm; 70~150rpm, Undefined
Alarm limit: EtCO2 High: (EtCO2 Low limit +1mmHg) ~ 150mmHg
EtCO2 Low: 0 ~ (EtCO2 High limit -1mmHg)
FiCO2 High: 0 ~ 150mmHg
AwRR High: (AwRR Low limit +1 rpm) ~ 120rpm
AwRR Low: 0 ~ (AwRR High limit -1 rpm)