EO-H Series

Ethylene Oxide Gas Sterilizer

Technical Specifications

PARAMETER

SPECIFICATION

Product name

Ethylene Oxide Sterilizer

Product form

Vertical

Heating method

Electric heating

Power

Volume

Chamber size

H80: 2.6KVA / 70L (Liner size: 585 x 275 x 422mm)

H120: 3.0KVA / 105L (Liner size:585 x 375 x 472mm)

H220: 3.3KVA / 195L (Liner size:700 x 505 x 552mm)

H330: 3.6KVA / 285L (Liner size:760 x 545 x 682mm)

Power

AC220V, 50Hz

Temperature display

3.5 inch LCD display

Material of Sterilization Room

304 anti-corrosion stainless steel

Outer box material

Carbon steel spray plastic

Temperature during sterilization

50℃±3℃

Working pressure

-60kpa

Highest anti-stress value

-140kpa

Humidity during sterilization

45%-80%

Selected gas

Pure ethylene oxide gas

Sterilization time

0-99 hours adjustable

Air Replacement

Automatic ventilation, 5 cycles of ventilation, every 15 minutes

Full-automatic control system

One-click start to finish, the whole process is completed

automatically

Sterilization record

The system can store 500 pots of sterilization records

Display status

Display operating status, temperature, humidity, pressure, sterilization time, residual gas treatment time, times, steam generator temperature.

Humidity Sensor

The built-in humidity sensor automatically detects the humidity in the cavity.

Pressure Sensor

The built-in pressure sensor automatically detects the pressure in the cavity

Temperature Sensor

The built-in temperature sensor automatically detects the temperature in the cavity, the accurate range is 50℃±3℃.

Vacuum system

Use oil-free vacuum pump.

Seal door

304 stainless steel aluminum material, thickness ≥ 2mm, pressure resistance 200Kpa.

Humidification System

Using pulse steam humidification

Residual gas treatment device

The filtration rate of the filter to the particle diameter ≥0.3μm is

not less than 99.5%

Vacuum rate

Under no-load conditions, the time for the sterilizer to be vacuumed from normal pressure to -60 kPa ≤ 30 min

Automatic dosing

Puncture technology built into the sterilization chamber.

Heating system

The explosion-proof electric heating device is used in the sterilization chamber

Multiple protection functions

With over-voltage and over-temperature protection functions, it is

more safe and reliable

Alarm system

Have high temperature alarm system, ultra low pressure alarm system