Products

AW-EA

Infant Warmer

Features

1. Microprocessor based servo controlled temperature system

2. The microprocessor-controlled unit allows for two modes: manual and automatic

3. Set temperature and skin temperature are displayed by LED

4. Alarm functions :

    Power failure, temperature deviation, temperature sensor failure, skin temperature failure, over temperature

5. A second thermal cut-out function for more safety

6. LED observation light adjustable,brightness adjustable

7. Adopts anti-explosion mirco crystal quartz infrared radiation tube

8. Four Transparent protector folded with 2 infusion seals

9. RS-232 connector, 1 tray

 

Specifications

Performance

Power supply

AC110/220V, 60/50Hz

Power input

900VA

Skin Temperature control range

32°C~38°C

Skin temperature control accuracy

≤0.5°C

Skin temperature sensor accuracy

±0.3°C

Mattress temperature uniformity

<2°C

Warm-up time(from 25°C)

<30min

Distance from heater to mattress

80cm

Mattress size

67cm*54cm

Operational Condition

Environment temperature

18°C~30°C

Environment relative humidity

30%~75%

Atmospheric pressure

700~1060hpa

Environment air velocity of flow

<0.3m/s

Transport and storage

Environment temperature

-40°C~+55°C

Environment relative humidity

95

Atmospheric Pressure

500~1060hpa

Package

Each unit is packed one carton

Carton Size

110*81*83cm

N.W./G.W

40KG/60KG

Standard configuration

1) Main Bod (including the radiant source,control system, infant bed,bracket)

2) I.V Pole

3) Skin temperature sensor

4) Tray

5) Mattress

6) Transparent protector

7) Castors

8) RS-232

Copyright © 2001-2024 ARI Medical Technology Co., Ltd. - ARI Technology Group Co., Ltd. All Rights Reserved.