Products

AR-323

Anesthesia Machine

Technical Specifications

PARAMETER

SPECIFICATION

Main Body

High-strength engineering plastic rack, light, beautiful and corrosion resistant

Scope of Application

Adult, child

Gas Source

O2: 0.28--0.6MPa;

N2O: 0.28--0.6MPa

Flow Meter

O2: 0.05--1.0L/min; 1.1--10L/min

N20: 0.05--1.0L/min; 1.1--10L/min

O2, N2O Linkage and N2O

Stopper

When using nitrous oxide, oxygen concentration>25%;

When the oxygen pressure < 0.2 MPa, the flow of nitrous oxide would be cut off.

Rapid Oxygen Flow

25--75L/min

Low Oxygen Pressure Alarm

There will be sound alarm when the oxygen pressure < 0.2MPa

Vaporizer

It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.

Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Respiratory Circuit

Working mode: all-close, semi-close, semi-open;

APL: 12.5kPa.

Respiratory Bellows

Bellows for adults, tidal volume range: 0--1500ml;

bellows for children, tidal volume range: 0--300ml.

Anyone can be chosen for application as required by the customer

VENTILATOR

Display Mode

High-definition 10.4” TFT color LCD Screen Display

Ventilation Mode

VCV, PCV, SIMV, MANUAL

Ventilation Function

PEEP, SIGH, IP, IRV, Standby

Ventilation Parameters

Tidal volume           20~1600ml

Rate                  2~99bpm

SIMV rate                   2~20bpm

I:E                  4:1--1:8

Inspiratory trigger pressure   -10~20cmH2O

PEEP                 2~20cmH2O

Pressure limitation     5~60cmH2O

SIGH         1.5 times the inspiratory time, 60-120 adjustable

Inspiratory platform       0~50%

Pressure control         3~50cmH2O

Pressure slope         1~10 gear

Parameters for Ventilation

Monitoring

Expired Tidal volume, Inhale tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, autonomous breathing rate, I/E, peak pressure of airway, average pressure, PEEP, inspiratory trigger pressure, Inspiratory platform, oxygen concentration, battery capacity, pressure-time waveform, flow-rate–time waveform, flow-volume loop, pressure-volume loop

SECURITY ALARM SYSTEM

Oxygen Concentration alarm

Upper limit setting range   21%--99%;

Low limit setting range     18%--80%.

Airway Pressure Alarm

Upper limit setting range   5--60cmH2O;

Low limit setting range     0--50cmH2O.

Minute Ventilation Volume Alarm

Upper limit setting range   1--20L/min;

Low limit setting range     0--19L/min.

Respiratory Rate

Upper limit setting range   10--99bPm

Low limit setting range     0--50bPm

Tidal Volume

Upper limit setting range   10--1500ml

Low limit setting range     0--1400ml

Sustained High-pressure

Alarm

It will give alarm when stress have consistently been higher than

25 cmH2O

Suffocation Alarm

Power alarm, Battery voltage alarm, Intubation off alarm

Power

AC 220V, 50Hz, Storage battery, UPS

Manipulators

(End-expiratory CO2 monitoring)

(Specific interface of waste anesthesia gas)

*Note: Items within brackets are optional.


Copyright © 2001-2024 ARI Medical Technology Co., Ltd. - ARI Technology Group Co., Ltd. All Rights Reserved.