Specifications

-Microcomputer control

-ESU-L2000 (LEEP) generator adopts LEEP which specializing in gynecological intrauterine event, increasing the accuracy, controllability, security of the surgery.

-Dealing with some kinds of pathological changes in obstetrics and gynecology, for example, cutting, electrocoagulation, cervical erosion, and some kinds of surgery that fit in peritoneoscope.

-Different electrodes have different effects in coagulation.

-An optional smoke evacuation system is available.

-Particular illuminating ESU pencil can increase the clarity of wound healing.

Technical Specifications

PARAMETER

SPECIFICATION

Power

220V±22V, 50Hz±1Hz

Operating Frequency

360460kHz

Power Rating

880VA±10

Five Working Modes

Monopolar Cut

a) Cut1 (pure cut): 1W~200W (Load 500Ω)

b) Cut2 (Blend1): 1W~150W (Load 500Ω)

c) Cut3 (Blend2): 1W~100W (Load 500Ω)

Monopolar Coag

d) Coagulation1 (Spray Coag): 1W~120W (Load 500Ω)

e) Coagulation2(Forced Coag): 1W~80W ( Load 500Ω)

Power Consumption

1100VA

SURGICAL SAMOKE EVACUATION

Blowing Rate

100 l/min @ 6mm tube

Noise

<60dB(A)

Start Control

Manus/Foot/Synchronize

Suction Adjust

10%~100%

Stop Delay Time

0~60s

AC Power

AC90~250V, 50Hz

Power Consumption

< 350W

Filter Characteristics

99.999% @ 0.1-0.2μm

Active Carbon

97% @ 1 year

Packing List

Mainframe                         1set

Electrosurgical pencil                3pieces

Illuminating electrosurgical pencil      2pieces

Electrosurgical pad                  10 pieces

Electrosurgical pad cable             1piece

Foot switch                         1set

Electrodes                          1case(10pieces)

Tip clean sponge                    1piece

Fuse                               3pieces

Power cord                         1piece

Operating manual                   1book