Products
-
- Operating Theatre Equipment
- OB/GYN Equipment
- Diagnostic Equipment
- Sterilizer (Autoclave)
- X-ray Series
- Laboratory Equipment
- Biological Microscope
- Spectrophotometer
- High Speed Centrifuge
- Low Speed Centrifuge
- High Speed Freeze Centrifuge
- Low Speed Freeze Centrifuge
- Special Centrifuge
- Thermostatic Bath
- Laboratory Shaker
- Laboratory Mixer
- Water Distiller
- Magnetic and Electric Stirrer
- Laboratory Incubator
- Dry Oven
- PH Meter
- Biological Safety Cabinet
- Clean Workbench
- Water Purifier
- Ultrasonic Cleaner
- Pipette
- Microtome
- Medical Hand Sink
- Medical Freezer
- Blood Bank Refrigerator
- Medical Pharmacy Refrigerator
- Analysis Balance
- Medical Destroyer
- In-Vitro Diagnostics
- Hemodialysis Equipment
- Hospital Furniture
- Walking Aids
- First-Aid Products
- Medical Endoscope
- Medical Human Model
- Ophthalmic Equipment
- Dental Equipment
- ENT Equipment
- Home Care Equipment
- Veterinary Equipment
- Medical Consumables
- Personal Protective Equipment
Features
-ASH-1800 is an economical and practical digital X-ray system, which can be widely used in radiology, orthopedics, health examination, emergency room, ICU, operating room and neonatology department. Fully meet the needs of digital radiation in hospital.
-40kW super power high frequency and high voltage generator and high heat capacity ball tube provide powerful guarantee for high quality X-ray output.
-Wireless flat panel detector, get rid of cable bondage.
-Wide nose angle, flexible positioning, hovering at any position within the angle range.
-The battery has a large energy storage capacity and can be exposed 1500 times in full charge state.
Product performance and parameters
Project |
Model |
Features and Performance |
High voltage generator |
BWG-50 |
External charging general power supply 220VAC, Lithium battery energy, full power can meet 200 times of exposure, Remote exposure controller > 50m, Tube Voltage: 40kV~150kV, Step Length: 1kV, Tube current:10mA~500mA, Load time range: 0.01s~6.3s, Current time product range: 0.1 mAs~250 mAs, Output nominal electric power: 40 kW (400 ma ,100 kv ,0.1 s), Inverter input supply voltage: 450 VDC/48VDC±10%, High voltage transformer structure: non-working frequency, Tube voltage: 40-125kv. |
X ray tube |
E7239X (CANON) |
Dual-focus 1.0mm/2.0mm, power 20/46KW, Maximum current 500mA maximum voltage> 125KVP, Heat storage capacity 140khu; 3200 rpm rotating anode. |
Collimator |
BRC-20 |
1.0mm Al/75kV; Manual control; LED visual field indicator light; 360° Adjustable rotation angle. |
Flat Panel Detector |
Mars1717V-PSI |
Cesium iodide amorphous silicon plate. Pixel matrix: >3072x3072; imaging area: 17x17 " Pixel size: 139μm, spatial resolution: 3.4lp/mm, AD conversion 16bits; Battery Autonomy 8hours;Full image time 5 seconds; AED mode |
Mobile frame |
|
Main wheel in front design; stable center of gravity; small turning radius; light motion control; easy to push by manpower; electromagnetic locking and fixation of cross arm during trolley motion; cross arm: front and back movement range is 250mm, up and down movement range is 1000mm; Max. Focus height is 191cm, Min. Focus height is 91cm; x-ray tube can rotate± 90°; head with locking mechanism can be locked at any angle. The column can be kneaded at ± 180°, and the electromagnetic brake with power loss can be locked and fixed. |
DR image processing Software |
|
DR diagnostic software: DICOM3.0 criteria, pre-selection tips for cast position; preset of image brightness / contrast, window width / window position; preset of organization equalization technology; preset of image enhancement and sharpening technology; image segmentation, transmission, printing, etc. |
Image Acquisition Station |
BWS-20 |
Minimum performance requirements for display: screen range: 17"; resolution :≥1280×1024, Minimum computer performance requirements: CPU≥i3 hard disk ≥500 G memory ≥4 G; |
Top panel
-State display: power display, 220 V voltage, inverter input voltage;
-USB interface: USB 3.0 interface, can be used to upgrade programs and copy materials, only for debugging mode
-Network port: used to connect network cable and communicate with external network
-Car armrest: push the armrest of the car, with an electric brake switch. When the switch is pressed, the brake is released, and the brake is locked when the switch is released; the rear steering wheel controls the steering movement of the machine by controlling the different forces exerted by two hands.
-Exposure handbrake: photographic exposure is carried out through the handbrake operation. The buttons on the top of the exposure handbrake have three positions: close, prepare and expose;
-Key switch: switch and operate the whole machine, "off" gear and "on" position
-Emergency stop button: the main switch of battery power supply. When the battery power supply is disconnected, the battery power supply is completely disconnected from the system, and all operations are prohibited.
Head Parts
-X-ray tube: X-ray is produced during photographic examination;
-Angle display: used to display the rotation angle of the ball tube and the beam limiter along the vertical axis;
-Handlebar: it is used to protect the head part from being hit. The other function is to operate the arm as a handrail, the arm lifting and stretching, and the column rotation;
-Brake control button: used to release the rack movement lock brake.