AR-7500A

High Frequency Digital Radiography system (DR)

Features

-50Kw high power, high frequency and high voltage generator and high thermal capacity tube

-Imported famous brand electronic components and chips; Strictly follow the GMP production quality management standard

-The original BGD sequential interconnection image acquisition system with the powerful image algorithm independently developed by the R&D team

-Capable of high KV photography

-Remote examination in compartments

-Remote PACS interconnection

-Remote maintenance and parameter calibration

-The image post-processing function is powerful



Product performance and parameters

Project

Model

Features and Performance

High Voltage Generator

(MOS Technology)

DTXR-50

Input: 110v/220v/380v,

Output power: 50kw,

Inverter frequency >300KHz,

Tube Current:10~500mA,

Current time product: 1mAs~630mA,

Tube Voltage: 40~125KV,

Can be directly connected to the network cable remote control for remote fault diagnosis and maintenance.

X-ray tube

(CANON)

E7242X

Dual-focus 0.6mm/1.5mm, power 20/46KW,

maximum current 500mA,

maximum voltage125KV,

Heat storage capacity 200KHU.

Flat Panel Detector

1800L

Strontium iodide amorphous silicon panel.

Pixel matrix > 3072 x 3072; Imaging area: 430x430mm" Pixel size: 140μm, Spatial resolution: 3.5Lp/mm, AD conversion 16bits; MTF:> 60% @1lp/mm> 20% @3lp/mmDQE:(@RQA5, 30μGy)>65% @0lp/mm >20% @3lp/mm Dynamic Range: >80dB

Imaging time: The output time of the original image is less than 3 seconds, and the post-loading software processing should be about 8 seconds.

Photograph Bed

SYC-2

Detector longitudinal motion range: 1000mm,

Table lateral movement 180 mm. Bed panel material: polycarbonate polymer panel,

Mechanical Motion Control: Manual electromagnetic lock. SID1000mm, Korean vibrating grid.

Tube stand

ZJ-1

The X-ray tube assembly moves up and down no less than 900mm;

Longitudinal movement is not less than 1800mm;

X-ray tube rotates along the horizontal axis ≥ ± 180 °

Collimator

XSQ-1

The total number of deviations between the sides of the ray radiation field and the respective sides of the light field does not exceed 3% of the distance from the field plane to the focal point. The sum of the deviations of the two axes should not exceed 4% of the indicated focal point to the detector plane.

Photograph Stand

SYJ-1

Detector activity: Vertical movement no less than 900mm;

SID1800mm, Korean vibrating grid

DR Image Processing

Software Workstation

DR

BGD Sequential Interconnection Acquisition Software,

DICOM3.0 standard, pre-selection of projection site; preset image brightness/contrast, window width/window position; organization equalization technology preset; image enhancement and sharpening technology preset; image segmentation, transmission, printing, etc.

Can be connected with PACS system.

Computer Host

DELL

2011

CPU 2GHz or more,

memory > 4G hard disk > 2000G, can store pictures in large capacity

Professional Diagnostic Display

ML2002

Size: 21.3"; Type: Monochrome; Mpixel:2MP; Resolution: 1600x1200;

Max Brightness: 1900cd/m2; Grayscale tone:16bit.

Remote Box

 

Can be remote controlled tube stand and photography frame, suitable for large-size chest X-ray examination

Operation display

 

Size: 19"; Brand:Dell