Products

AR-300

Digital Mammography System

Standard Configuration

No.

Configuration

Quantity

1

High Frequency High Voltage Generator

1

2

X-ray Detector

1

3

X-ray Tube

1

4

Exposure Control Box and Exposure Footswitch

1

5

Quadruple Foot Switch

1

6

Image Acquisition and Processing System 

1

7

24cm×30cm Compression Paddle

1

8

Acquisition Workstation (host computer, software,

2MP grayscale medical grade display)

1

Optional Configuration

No.

Configuration

Quantity

1

Diagnostic Workstation (graphics workstation, software,

2×5MP grayscale medical grade display)

1

2

Localization Paddle:18cm×24 cm

1

3

15cm×15cm Compression Paddle

1

4

Spot Contact Compression Paddle: (diameter) 8cm

1

5

Magnification Compression Paddle: 18cm x 24cm

1

6

Magnification Platform

1

7

Localization Crosshair Assemblies

1

Product Specification

l  General Operating Conditions

Temperature range: 10℃ to 40℃

Relative humidity range: 30% to 75% non-condensing

l  Storage Environment

Storage Temperature Range: -20 to 50

Relative Humidity Range: 20% to 95% non-condensing

l  Main Technical Parameters

(I) Generator

1.1. Inverting frequency of generator: ≥28kHz

1.2. Maximum Output: 5 kW

1.3. kV Range: 20 kV to 40 kV

1.4. mAs Range: 0.1 mAs to 630 mAs

1.5. Maximum mA: 166 mA

(II) X-Ray tube:

2.1. Anode Type: Molybdenum

2.2. Anode Design: Bi-angular

2.3. Focal Spot Size: 0.1 mm (Small), 0.3 mm (Large)

2.4. Anode Speed: ≥9600 RPM

2.5. Heat Capacity: 300KHU (225 kJoules)

2.6. Heat dissipation power: 715W

2.7. Tube voltage: 35KV

2.8. Anode heat storage capacity: ≥500kHu

2.9. Collimator Type: electric, automatic adjustment

2.10. Collimator feature: automatic window adjustment

2.11. Supports horizontal and vertical collimator blades

(III) Digital Image detector:

3.1. Detector: self-develop and self-produced detector

3.2. Type: Photon-counting Direct Capture Technology

3.3. X-Ray Absorption Material: monocrystalline silicon

3.4. High-resolution acquisition matrix: ≥3000×4000

3.5. Pixel size: ≤70μm

3.6. Output Image: ≥16 bit image data

3.7. Limiting Spatial Resolution: 7 lp/mm

3.8. Without grid

3.9. Detector operating temperature range: 10°C-40°C;

Operating humidity range: 30%-75%

3.10. Effective Image Area: 21 cm x 28 cm

3.11. Magnification factor:1.5

3.12. Ghosting Coefficient: <0.1

(IV) C-Arm Assembly

4.1. Vertical Travel: 76 cm to 134 cm motorized

4.2. Rotation: +200° to – 200°

4.3. Rotation type: isocenter rotation

4.4. SID: 65 cm

4.5. C-arm control panels: up-down and rotation movement control

4.6. LCD Parameter display screen: thickness, angle, and pressure

4.7. Compression fine-tuning knobs

4.8. Manual pressure release handle

4.9. Emergency stop buttons (available on both the mammographic stand and workbench)

4.10. Compression one-key release bottom on workbench

4.11. Patient handrail

4.12. Patient face shield

4.13. Can be adapted to foot pedals

4.14. One-key positioning function: one-key switch to the next position, one-key mirror image, one-key reset

(V) Compression Device

5.1. Compression type: Electrically powered

5.2. Compression Control: Manual, electric, foot pedal control

5.3. Compression Mode: Intelligent flexible compression, manual compression, manual fine-tuning

5.4. Compress release: automatic release (which can be enabled and disabled by software setting), one-key release, manual release control

5.5. Compression force (electric): 0-200N

5.6. Maximum compression travel: 250mm

5.7. Compression paddle size: 24cm×30cm, 15cm×15cm(optional config), 8cm (diameter) Spot Contact(optional config), 18cmx24cm(optional config), Localization Paddle: 18cm×24 cm (optional config)

5.8. Automatically recognize the paddle type

(VI) Acquisition Workstation

6.1. CPU: Multi-Core Intel Based CPU

6.2. Operating System: Windows/Linux

6.3. Memory: ≥16 GB

6.4. Hard Drive: ≥512 GB

6.5. Monitor Size: ≥22 inch

6.6. Monitor display resolution: ≥1280×1024

6.7. Image Display:

2MP grayscale medical grade display (standard)

5MP grayscale medical grade display (optional)

6.8. DICOM Services: Worklist, Print, Storage, Query, Retrieve

6.9. Local Image Store Capacity: Approximately 11,000 screening

6.10. Compatible with PACS/RIS/HIS

6.11. Basic image post-processing functions: overall image resizing, image moving, window width/ level adjustment, local image amplification, image color reversal, image flipping/ rotation, and text input

6.12. Image display supports multiple layouts

6.13. One-click restoration of window width and window level settings

6.14. One-click multi-image synchronous image processing and display

6.15. Exposure Modes:

Manual (Operator selects all parameters)

Auto-Time (System selects mAs, Operator selects filter and kV)

Auto-kV (System selects kV and mAs, Operator selects filter)

Copyright © 2001-2024 ARI Medical Technology Co., Ltd. - ARI Technology Group Co., Ltd. All Rights Reserved.