Products
-
- Operating Theatre Equipment
- OB/GYN Equipment
- Diagnostic Equipment
- Sterilizer (Autoclave)
- X-ray Series
- Laboratory Equipment
- Biological Microscope
- Spectrophotometer
- High Speed Centrifuge
- Low Speed Centrifuge
- High Speed Freeze Centrifuge
- Low Speed Freeze Centrifuge
- Special Centrifuge
- Thermostatic Bath
- Laboratory Shaker
- Laboratory Mixer
- Water Distiller
- Magnetic and Electric Stirrer
- Laboratory Incubator
- Dry Oven
- PH Meter
- Biological Safety Cabinet
- Clean Workbench
- Water Purifier
- Ultrasonic Cleaner
- Pipette
- Microtome
- Medical Hand Sink
- Medical Freezer
- Blood Bank Refrigerator
- Medical Pharmacy Refrigerator
- Analysis Balance
- Medical Destroyer
- In-Vitro Diagnostics
- Hemodialysis Equipment
- Hospital Furniture
- Walking Aids
- First-Aid Products
- Medical Endoscope
- Medical Human Model
- Ophthalmic Equipment
- Dental Equipment
- ENT Equipment
- Home Care Equipment
- Veterinary Equipment
- Medical Consumables
- Personal Protective Equipment
Standard Configuration
No. |
Configuration |
Quantity |
1 |
High Frequency High Voltage Generator |
1 |
2 |
X-ray Detector |
1 |
3 |
X-ray Tube |
1 |
4 |
Exposure Control Box and Exposure Footswitch |
1 |
5 |
Quadruple Foot Switch |
1 |
6 |
Image Acquisition and Processing System |
1 |
7 |
24cm×30cm Compression Paddle |
1 |
8 |
Acquisition Workstation (host computer, software, 2MP grayscale medical grade display) |
1 |
Optional Configuration
No. |
Configuration |
Quantity |
1 |
Diagnostic Workstation (graphics workstation, software, 2×5MP grayscale medical grade display) |
1 |
2 |
Localization Paddle:18cm×24 cm |
1 |
3 |
15cm×15cm Compression Paddle |
1 |
4 |
Spot Contact Compression Paddle: (diameter) 8cm |
1 |
5 |
Magnification Compression Paddle: 18cm x 24cm |
1 |
6 |
Magnification Platform |
1 |
7 |
Localization Crosshair Assemblies |
1 |
Product Specification
l General Operating Conditions
Temperature range: 10℃ to 40℃
Relative humidity range: 30% to 75% non-condensing
l Storage Environment
Storage Temperature Range: -20℃ to 50℃
Relative Humidity Range: 20% to 95% non-condensing
l Main Technical Parameters
(I) Generator
1.1. Inverting frequency of generator: ≥28kHz
1.2. Maximum Output: 5 kW
1.3. kV Range: 20 kV to 40 kV
1.4. mAs Range: 0.1 mAs to 630 mAs
1.5. Maximum mA: 166 mA
(II) X-Ray tube:
2.1. Anode Type: Molybdenum
2.2. Anode Design: Bi-angular
2.3. Focal Spot Size: 0.1 mm (Small), 0.3 mm (Large)
2.4. Anode Speed: ≥9600 RPM
2.5. Heat Capacity: 300KHU (225 kJoules)
2.6. Heat dissipation power: 715W
2.7. Tube voltage: 35KV
2.8. Anode heat storage capacity: ≥500kHu
2.9. Collimator Type: electric, automatic adjustment
2.10. Collimator feature: automatic window adjustment
2.11. Supports horizontal and vertical collimator blades
(III) Digital Image detector:
3.1. Detector: self-develop and self-produced detector
3.2. Type: Photon-counting Direct Capture Technology
3.3. X-Ray Absorption Material: monocrystalline silicon
3.4. High-resolution acquisition matrix: ≥3000×4000
3.5. Pixel size: ≤70μm
3.6. Output Image: ≥16 bit image data
3.7. Limiting Spatial Resolution: 7 lp/mm
3.8. Without grid
3.9. Detector operating temperature range: 10°C-40°C;
Operating humidity range: 30%-75%
3.10. Effective Image Area: 21 cm x 28 cm
3.11. Magnification factor:1.5
3.12. Ghosting Coefficient: <0.1
(IV) C-Arm Assembly
4.1. Vertical Travel: 76 cm to 134 cm motorized
4.2. Rotation: +200° to – 200°
4.3. Rotation type: isocenter rotation
4.4. SID: 65 cm
4.5. C-arm control panels: up-down and rotation movement control
4.6. LCD Parameter display screen: thickness, angle, and pressure
4.7. Compression fine-tuning knobs
4.8. Manual pressure release handle
4.9. Emergency stop buttons (available on both the mammographic stand and workbench)
4.10. Compression one-key release bottom on workbench
4.11. Patient handrail
4.12. Patient face shield
4.13. Can be adapted to foot pedals
4.14. One-key positioning function: one-key switch to the next position, one-key mirror image, one-key reset
(V) Compression Device
5.1. Compression type: Electrically powered
5.2. Compression Control: Manual, electric, foot pedal control
5.3. Compression Mode: Intelligent flexible compression, manual compression, manual fine-tuning
5.4. Compress release: automatic release (which can be enabled and disabled by software setting), one-key release, manual release control
5.5. Compression force (electric): 0-200N
5.6. Maximum compression travel: 250mm
5.7. Compression paddle size: 24cm×30cm, 15cm×15cm(optional config), 8cm (diameter) Spot Contact(optional config), 18cmx24cm(optional config), Localization Paddle: 18cm×24 cm (optional config)
5.8. Automatically recognize the paddle type
(VI) Acquisition Workstation
6.1. CPU: Multi-Core Intel Based CPU
6.2. Operating System: Windows/Linux
6.3. Memory: ≥16 GB
6.4. Hard Drive: ≥512 GB
6.5. Monitor Size: ≥22 inch
6.6. Monitor display resolution: ≥1280×1024
6.7. Image Display:
2MP grayscale medical grade display (standard)
5MP grayscale medical grade display (optional)
6.8. DICOM Services: Worklist, Print, Storage, Query, Retrieve
6.9. Local Image Store Capacity: Approximately 11,000 screening
6.10. Compatible with PACS/RIS/HIS
6.11. Basic image post-processing functions: overall image resizing, image moving, window width/ level adjustment, local image amplification, image color reversal, image flipping/ rotation, and text input
6.12. Image display supports multiple layouts
6.13. One-click restoration of window width and window level settings
6.14. One-click multi-image synchronous image processing and display
6.15. Exposure Modes:
Manual (Operator selects all parameters)
Auto-Time (System selects mAs, Operator selects filter and kV)
Auto-kV (System selects kV and mAs, Operator selects filter)